This document provides instructions for 

Procedure

for troubleshooting, checking & restarting Optifacts programs and interfaces. 

Procedure

Types of Interfaces

Optifacts programs

Lab Equipment

Scanners

Steps to take:

Insure the demons are running in menu 10-2

From the Optifacts main menu:

Type 10 - Management Functions and press enter.

Type 2 - Check/Restart Optifacts Programs and press enter.

Options:

  • R - Restart
  • K - Stop Interface
  • F - Refresh the screen
  • L - View the log files

If the problem demon is running, try resetting it.

Image Added

  • Push the ( K ) key to stop the demon.
  • Then, press enter through the remaining demons.
  • Select 2 again and you should see the following:

Image Added

  • Select R to restart the selected demon.
  • Press enter through the remaining demons again and the demons should be running.

Image Added

Examples of various types of programs and interfaces

Image Added     Image Added

Try communicating with the piece of equipment again. If it still fails, check the cabling and connectors.

Cabling

  • Ensure that the connector is securely snapped to the port on the Rocket or Digi Port.
  • Ensure that the connector is securely snapped to the connector on the equipment.
  • Make sure that no rust has developed on any of the connectors.

Connectors