  1. ART-M request data for a job.
  2. LMS responds with data for both eyes
  3. ART-M will upload SVAL, AVAL, KPRVA, KPRVM, PIND, GRPVA, GPRVM, _BLKDM, _BLKRD and _BLKCEHI.  The data for the RIGHT eye is what was actually blocked; the data for the left eye is a mirror of what was received from the LMS.
  4. ART-M will request data for the same job a second time; this is used for the left eye blocking.  Therefore, the data uploaded for the right eye must be re-sent to the blocker, but the left eye data should be that as calculated by the LMS
  5. ART-M will upload SVAL, AVAL, KPRVA, KPRVM, PIND, GPRVA, GPRVM, _BLKDM, _BLKRD, and _BLKCEHI with data for the LEFT eye; the right eye data is also present, and is a mirror of what was received from the LMS in step 4.

Image Added

In Rx-Universe, the device session for the ART-M should be set up with 2.00 degrees of prism to be sent to the generator (with the remainder blocked), and a maximum blocker prism of 4.00.  (Note that the generator prism is a guideline; the ART-M will calculate the actual amount of prism to be both blocked and generated).  The checkbox indicating the device is an ART blocker must also be checked.

...